P2000 Specifications
Maximum Input Voltage: DC 2,000 V: INPUT H - INPUT L, Maximum Rated Voltage
[Maximum Rated Voltage Between ground] 1,000 V (Measurement Category IV) Expected Transient Voltage 12,000 V, 2,000 V (Measurement Category III) Expected Transient Voltage 1,5,000 V
Input resistance: 20 MΩ ± 5.0% (INPUT H - INPUT L intervals, with OUTPUT terminal open).
Overload Protection: DC 2,200 V/AC 2,200 V 1 minute (INPUT H – INPUT L interval), DC 600 V/AC 600 V 1 minute (OUTPUT H – OUTPUT L intervals).
[OUTPUT Terminal] 0.2 inch (4 mm) banana terminal.
Operating Temperature Range: -5°F to 145°F (-25°C to 65°C).
Operating Humidity Range: -5°F to 104°F (-25°C to 40°C): 80% RH or less (non-condensation), 40°C to 65°C: 40°C 80% RH or lower to 65°C 25% RH (do not condensation),
Storage temperature and humidity range: -30°C - 70°C, 90% RH or lower (do not condensation)
[Compliance Standards] EN 61010
Accessories: Connection cable L4943 x 1, Strap buckle x 2, Strap x 1, Carrying case C0205 x 1, Instruction manual x 1, Usage Precautions x 1.
[Model Name] Combination accuracy specification
[DT4261] Range 600.0 V: Combination Accuracy: ±0.5% rdg ±0.2 V, Range 2000 V: Combination Accuracy: ±0.5% rdg ±5 V
[CM4141-50, CM4371-50, CM4373-50, CM4375-50) Range 600.0 V: Combination Accuracy: ±1.0% rdg ±0.3 V, Range 2000 V: Combination Accuracy: ±1.0% rdg ±3 V
Accessories: [Accessories] Connection cable L4943 x 1, Strap buckle x 2, Strap x 1, Carrying case C0205 x 1, Instruction manual x 1, Usage Precautions x 1.